data.edit.corr.aeth_schmid_coen
data.edit.corr.aeth_schmid_coen calculates absorption coefficients from Aetholemeter data. It requires the attenuation (from data.edit.corr.aeth_calc_abs), the attenuation, scattering, and the filter segmentation for the instrument in the $INSTRUMENT_filter segments.
data.edit.corr.aeth_schmid_coen can operate as either an independent filter (provided the correct input is given) or as part of the correction chain
Command Line Usage
data.edit.corr.aeth_schmid_coen [--input=A11] [--neph=S11] [--m=0.74] [--c=3.29E-4] [--d=0.564] [--cref-WL=...]
Arguments
--input=A11
Specify what instrument or record base to use as the input, instead of trying to match what's available. Required if the input stream is ambiguous.
--neph=S11
Specify what instrument or record base to use as the neph data, instead of trying to match what's available. Required if the input stream is ambiguous.
--m=0.74
Set the value for “m” defaulting to 0.74.
--c=3.29E-4
Set the value for “c” defaulting to 3.29E-4.
--d=0.564
Set the value for “d” defaulting to 0.564.
--cref-WL=2.1
Set the value for Cref for the wavelength WL (in nm), defaulting to:
Wavelength | Cref |
---|---|
370 | 1.813 |
470 | 2.073 |
520 | 2.076 |
590 | 2.104 |
660 | 2.182 |
880 | 2.226 |
950 | 2.199 |
Example Usage
data.get mlo A81a,S11a 2010:1 1d | data.edit.corr.aeth_schmid_coen